6. SETTING AND CALIBRATION

6.1 Switch of range

6.1.1 Setting of range switch mode

ENT
ed.
Description of setting
MR: Select a desired range on this screen.
RR: Select a desired range according to the
remote range switch contact input.
AR: Automatically switched from Range 1 to
Range 2 when the measured concentration
exceeds 90% of Range 1.
Automatically switched from Range 2 to
Range 1 when the measured concentration
becomes less than 80% of Range 1.
* Operation set for each Ch only can be
performed.
MODE
key.
ENT
key.
ENT
Measurement Mode Screen
User Mode
Switch Ranges
Calibration Parameters
Setting of Auto Calibration
Setting of Auto Zero Calibration
Parameter Setting
-
Switch Ranges
-
Ch1
MR
CO
2
Ch2
AR
CO
-
-
Switch Ranges
Ch1
MR
CO
2
Ch2
AR
CO
or previous screen
22
MODE
Select an item
with UP/DOWN and ENT
Back with ESC
ENT
Select Ch No.
with UP/DOWN and ENT
Back with ESC
Range1
0–10.00
Range2
0–20.00
Range1
0–500.0
Range2
0–2000
ENT
(
)
Select method of
switch ranges
with UP/DOWN and ENT
Back with ESC
Range1
0–10.00
Range2
0–20.00
Range1
0–500.0
Range2
0–2000
ENT
Range switch
INZ-TN2ZPAH-E
vol%
vol%
ppm
ppm
vol%
vol%
ppm
ppm