6. ZERO CALIBRATION AND SETTING

Select "Zero Calibration" from the "Menu" screen and press the
Menu

Zero Calibration

Span Calibration
Alarm Log
Alarm Setting
Output Hold

6.1.1 Preparation

AUTION
Parts name
Calibration gas cell
Receiver/transmitter cable (for cali-
bration)
Receiver/transmitter cable
Zero gas (N
)
2
Pressure adjuster (regulator)
Pipe (Teflon tube, etc)
Joint (Rc1/4)
Flow meter
Thermometer
Pressure gauge
Others (joint, tools, etc)
(1) Turn OFF the power.
(2) Remove the receiver box and the transmitter box using the hexagonal wrench. Perform removing when
the provided equipment is stopped. Never remove boxes while the equipment is operated.
Otherwise, hot temperature gas may blow out.
6.1 Zero calibration
2006-04-01
12:00:00
Parameter Setting
Display/Output
Input Signal
AO Select
Password
*****
• The laser beam is the invisible infrared light. Do not watch the laser
beam directly or scattering light.
• Do not watch the laser beam directly with the optical measuring
device. Otherwise, it may cause serious damage to your eyes.
Quantity
1
1
1
1
1
Several m
2
1
1
1
- 34 -
key.
NT
Zero Calibration
Component
Cell Length
Cal.Gas Value
Meas.
NT
LastCal.Day
2006/09/01
Remarks
To be ordered separately
To be ordered separately
To be ordered separately
To be ordered separately
To be ordered separately
φ 10 × 8 or more
For calibration gas cell
2L/min or more
4 to 20mA output, for temperature correc-
tion
4 to 20mA output, for pressure correction
2007-4-1
12:00:00
HC ℓ
1000
mm
0000
ppm
00.12
ppm
ZeroCal.Start
Cal.Finish
INZ-TN1ZSS-E