6. SETTING AND CALIBRATION

6.1 Switch of range

6.1.1 Setting of range switch mode

MODE
ENT
ed.
Description of setting
MR: Select a desired range on this screen.
RR: Select a desired range according to
the remote range switch contact input.
AR: Automatically switched from Range 1
to Range 2 when the measured concen-
tration exceeds 90% of Range 1.
Automatically switched from Range 2
to Range 1 when the measured concen-
tration becomes less than 80% of
Range 1.
* Operation set for each Ch only can be
performed.
key.
ENT
key.
ENT
36
Measurement Mode Screen
MODE
User Mode
Select an item
with UP/DOWN and ENT
Back with ESC
Switch Ranges
Calibration Parameters
Alarm Setting
Setting of Auto Calibration
Setting of Auto Zero Calibration
Setting of Peak Alarm
Parameter Setting
ENT
Switch Ranges
Select Ch No.
with UP/DOWN and ENT
Back with ESC
Ch1
Range1
MR
NOx
Range2
Ch2
Range1
AR
O
Range2
2
Switch Ranges
Select method of
switch ranges
with UP/DOWN and ENT
Back with ESC
Ch1
Range1
MR
NOx
Range2
Ch2
Range1
AR
O
Range2
2
ENT
Range switch
or previous screen
0–10.00
ppm
0–100.0
ppm
0–10.00
vol%
0–25.00
vol%
ENT
(
)
0–10.00
ppm
0–100.0
ppm
0–10.00
vol%
0–25.00
vol%
INZ-TN2ZPG-E