Calibration / Deskew Procedure - Keysight InfiniiMax III+ Series User Manual

Probe
Table of Contents

Advertisement

Keysight InfiniiMax III+ Series Probes
User's Guide
4 Calibration / Deskew
Procedure
The Calibration/Deskew procedure described in this chapter is applicable to both
N2830/1/2A and N7000/1/2/3A InfiniiMax III+ probes.
The probe calibration and deskew is a guided procedure that you start from the
oscilloscope's Probe Calibration dialog box. Depending on the oscilloscope model,
you will be instructed to connect the probe head to either of the following
oscilloscope outputs:
• Front-panel Probe Comp terminals
• Aux Out BNC connector or Cal Out connector using the N5443A
Calibration/Deskew Fixture with 50W termination.
The N5443A fixture promotes the properly position the probe head during the procedure. If
N OT E
you prefer, you can remove the plastic holder from the fixture by removing the four screws on
the bottom side of the holder.
Always wear an ESD wrist strap when working with active probes. Not doing so can
CAUTION
result in the probe becoming permanently damaged.
85

Advertisement

Table of Contents
loading

This manual is also suitable for:

N2830aN2831aN7001aN7003aN7002aN7000a ... Show all

Table of Contents