Description Art. no. Designation Base of cart Main power switch. Press to turn the system ON. PV010 Aesculap DSM (Digital Surgical Microscope) Press again to turn the system OFF PV022 Software module DUV 400 Product label AC input PV024 Software module DICOM...
To ensure optimal image quality, only use product with 3D glasses sup- Aesculap DSM (Digital Surgical Microscope) PV010 plied by Aesculap. Contact Aesculap if additional pairs are required. The Aesculap DSM (Digital Surgical Microscope) generates a magnified 3D ► Do not use polarized prescription glasses as they will distort the 3D view of the surgical field.
Product description To prevent fire or shock hazard, do not expose system to rain or mois- ture. The Aesculap DSM is a freestanding, digital surgical microscope in a look- over configuration. The DSM does not utilize traditional microscope bin- Ambient conditions...
DSM head 3.2.2 Handle buttons (default setting) The DSM is designed to be controlled primarily through the buttons on the The DSM head contains two sets of optics and two high-resolution sensors handles of the sides of the head. The default layout of the handles is shown to convert the analog signal to a digital video stream.
► 3.3.2 Cart LED indicators Connect the display to the HDMI output port c at the back of the cart, see Fig. 5. There are three LED indicators near the base of the cart that display the status of the DSM, system computer (EPU), and robotic arm components. ►...
► Touchscreen Change the batteries of the footswitch when the battery indicator flashes orange, see Chapter 5.12. The system can be controlled by using the touchscreen. The touchscreen is 2D only and not sterile. 3.7.3 Backup USB cable ► Position the touchscreen for quick and convenient access. Footswitch PV014 The system can be controlled by using the wireless footswitch.
3.11 Additional components Working with the product The DSM system comes with additional components to help configure the system to work best in the operating room. Risk of injury and/or product malfunction due to incorrect operation of the electromedical system! Art.
IEC/DIN EN 60601-1 or applicable national standards. ► Please address the B. Braun/Aesculap partner or Aesculap Technical Service with any inquiries in this respect; for a contact address, see Chapter 10. ►...
b = 2,03 m c = 1,47 m Fig. 13 Legend Fig. 11 Sliders of the yaw coupler Legend ► Unlock the sliders a and rotate the DSM manually into the desired posi- Top of the screen tion. Max. monitor height (top of the screen) The available coupler positions are illustrated below.
► Starting a case Press “Yes” to continue saving the media information. – or – Once the DSM arm has been draped and all pre-operative checks are com- plete, a new case can be started. ► Press “No” to permanently delete all case media (this will require con- firmation).
5.9.2 Copy media to drive 5.12 Changing footswitch batteries Battery type: 3 × IEC- LR14 (Type C – “Baby”) ► Shut down DSM, see Chapter 5.11. ► Open battery housing: Rotate both lockers near battery housing by 90°. While doing so, lightly press down the battery cover. ►...
DSM setup 6.2.2 Rebalance scope Moving the robotic arm to its limits can affect the DSM’s balance. To cor- The “Setup” panel in the home screen provides the following functions: ■ rect the presence of any drift, the DSM can be rebalanced. Move the robotic arm to a preset position ■...
Fluorescence controls Touchscreen robot control Note Fluorescence module (DUV 400) requires a license for use with the DSM sys- tem. DUV 400: The integrated DUV 400 fluorescence module of the DSM is used to visualize near-UV fluorescent areas in the surgical field. The mod- ule has been designed for excitation in the wavelength range from 390 nm to 420 nm and for fluorescence visualization in the wavelength range greater than 510 nm.
The following QAB functions are available: State Icon Description Function Description Footswitch Directional Displayed when moving the arm Move with Restrictions with the footswitch and arm reaches Viewport Configure the format for multiple inputs on screen. a directional limit Select between primary input, secondary input, split screen, or picture-in-picture.
Commonly found buttons in the settings panels include: DSM Settings The “DSM Settings” drop-down list has the following options: Icon Description Option Description Exit panel without saving changes made to current panel. Auto-Hide The DSM control values (Zoom, Focus, and White Light) (default) will be displayed over the live surgery screen for ~5 sec- Reset the current panel to its default settings.
White Balance – Squeezed SxS (side by side): checking the checkbox reduces resolu- tion of saved media to 1920 × 1080 (default: unchecked). Select Every DSM is shipped with pre-configured white balance settings. this option when smaller file sizes and/or the ability to edit with 3rd Note party editing software are required.
► Swap 3D Exit out of the DSM software onto the Windows Desktop, see Chapter 5.11. “Swap 3D” switches the left and right eye polarity and can be used to help determine if the monitor is using the correct 3D polarity setting. ►...
6.8.11 Footswitch More ► In the Footswitch panel the footswitch can be paired with the DSM Press “More” to open a separate panel with two pages. and the layout and function of the footswitch can be configured. Page 1 contains system information, including Computer Name, Mac See below for a list and description of each button’s function.
Page 23
Damage to the product due to inappropriate clean- ing/disinfecting agents and/or excessive tempera- tures! CAUTION ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which are approved for plastic material and high- grade steel, do not attack softeners (e.g. in silicone). ►...
Page 24
Troubleshooting list Malfunctions Error messages 9.2.1 Startup errors Malfunction Remedy Error message Remedy Colors look wrong Perform a white balance of the DSM head in the software settings. Already initialized Restart the system. If that does not help, contact the 3D looks inverted Switch the 3D polarity in the software settings.
Page 25
Aesculap Inc. ► Detailed information concerning the disposal of the product is avail- Attn. Aesculap Technical Services able through the national B. Braun/Aesculap agency, see Chapter 10. 615 Lambert Pointe Drive Hazelwood 13. Distributor in the US/Contact in Canada MO, 63042 USA...
Page 26
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com ® Aesculap – a B. Braun brand SOP-AIC-5001887 (TA015615) 2019-02...