5.3.2 Setting Example for Numeric-Type Data: Flow rate span

This example describes the setting of the flow rate span for the numeric-type parameter B23: Flow Span from
100 l/min to 120 l/min.
Setting mode
Start:
Major item
Major Item
parameter
Parameter
Search Mode
Major item
Sub-item
Parameter
Sub-item
Search Mode
parameter
Major item
Sub-item
selection (B)
Parameter
Replacement Mode
<5. BASIC OPERATING PROCEDURES
Setting mode
P:Protect
B:Easy Setup
C:Basic Setup
B:Easy Setup
50:Auto Zero Exe
10:Language
20:Flow Damping
INC
X4
B:Easy Setup
22:Base Time Unit
23:Flow Span
24:Flow Decimal Pnt
Selection of the appropriate parameter
SET
ESC
SET
B23:Flow Span
100 l/min
000100. l/min
Rng:0.00001 → 32000
/SHIFT
DEC
INC
ESC
SET
DEC/SHIFT
INC
B23:Flow Span
100 l/min
0000100. l/min
Rng:0.00001
32000
DEC /SHIFT
X4
INC
X2
SET
F0505-1.EPS
(USING THE DISPLAY UNIT)
Setting Mode Condition
Touch the SET
switch to access B: Easy Setup.
NOTE
The ▲ and ▼ symbols to the left of the
parameters indicate that additional setting items
to those being currently displayed may also be
INC
selected. Use the
these items.
Sub-Item Parameter Search Mode has been
accessed in this screen.
Touch the
INC
switch to move the cursor to B23:
Flow Span.
The cursor has been moved to B23: Flow Span in
this screen. (Sub-item selection screen (B))
Touch the
switch to access Parameter
SET
Replacement Mode.
Once Parameter Replacement Mode has been
selected, the digit that can be replaced will be
flashed on and off. When in this condition, confirm
the relevant setting range as displayed at the
bottom of the screen and then set the parameter as
required.
In this example, the parameter will be set to "120 l/min."
NOTE
When setting a new value, use the DEC /SHIFT
switch to move from digit to digit, and use the
INC
switch to cycle through values for each
individual digit. In addition to digit, it is also
possible to select a decimal point, and this allows
the position of the decimal point to be changed.
Modify the value to "120 l/min" as follows:
Touch the DEC /SHIFT switch to move the cursor
to the position for multiples of 10. Then, touch the
switch to change the value at this position
INC
from "0" to "2."
When the value of "120" has been setup, touch the
switch.
SET
5-7
>
switch to cycle through
IM 01E30D01-01EN