Wiring Procedure - YOKOGAWA ADMAG AXR User Manual

Two-wire magnetic flowmeter integral flowmeter
Table of Contents

Advertisement

4.6

Wiring Procedure

(1) Power supply and Current output
This instrument uses the same two wires for both,
the signal and power supply. A DC power supply is
required in a transmission loop.
The allowable power supply voltage is described
below.
• 14.7 to 42V DC for general-purpose use and
explosion proof type
• 14.7 to 32V DC for lightning Protector (optional
code A)
Wire the AXR according to Figure 4.6.1 or Figure
4.6.2.
WARNING
Before wiring with external instruments, be sure
to turn off the external instruments.
IMPORTANT
This instrument normally outputs a current of 12
mA for several seconds just after turning on the
power.
Warm up this instrument for at least 30 minutes.
Perform flow measurement 30 minutes after
turning on the power.
SUPPLY
DO
Figure 4.6.1 Electric Cable Wiring (General -
purpose Use/Explosion Proof Type
except TIIS)
SUPPLY
DO
Figure 4.6.2 Electric Cable Wiring (TIIS Explosion
• Connecting with the commercial AXR
• Do not connect power supply with reversed
Supply voltage means the voltage necessary
to provide between the power terminals of the
magnetic flowmeter.
Read Table 4.6.2 to Table 4.6.4 for wiring examples
in case of general - purpose use/explosion proof
type except TIIS.
Read Table 4.6.5 to Table 4.6.7 for wiring examples
in case of TIIS explosion proof type.
The ADMAG AXR can be connected with almost
all distributors, signal conditioner cards, and I/O
modules except certain devices.
Read the Table 4.6.1 for Yokogawa's devices,
Power supply
choose an appropriate connecting device and the
cable
corresponding length of cable.
F0414.ai
For devices other than Table 4.6.1, decide on
the connection by reading the supply voltage
specifications and the description of wiring
examples as shown in Table 4.6.2 to Table 4.6.7.
<4. WIRING>
Proof Type)
IMPORTANT
power supply will damage the flowmeter.
Be sure to use the DC power supply in the
predetermined range.
polarities.
SUPPLY + terminal: connect +
SUPPLY - terminal: connect -
NOTE
4-7
Power supply
cable
F0415.ai
IM 01E30D01-01EN