YOKOGAWA ADMAG AXR User Manual page 142

Two-wire magnetic flowmeter integral flowmeter
Table of Contents

Advertisement

This section describes the procedure for zero adjustment using the display unit switches. (For more details
regarding setting methods using these switches, read Chapter 5.)
The parameters for zero adjustment are B50/M10: Auto Zero Exe (and either of these can be used to carry
out this procedure). For more details regarding these parameters, read Chapter 6.
The parameter M10: Auto Zero Exe will be used in the following description.
Start:
Major Item
Parameter
Search Mode
Setting Mode
Major item
parameter
Major item
Sub-item
Sub-item
Parameter
selection (D)
Search Mode
Parameter
Replacement
(execute) Mode
Setting mode
K:Diagnosis
M:Adjustment
N:Test
SET
M:Adjustment
21:Adjustment 20mA
10:Auto Zero Exe
11:Autozero Time
SET
M10:Auto Zero Exe
No Execution
No Execution
Execution
INC
M10:Auto Zero Exe
No Execution
Execution
No Execution
SET
M10:Auto Zero Exe
Execution
SET
<10. ACTUAL OPERATION>
Once in Setting Mode, use the
move the cursor to M: Adjustment.
Touch the
switch to access Sub-item
SET
Parameter Search Mode.
Upon selection of M: Adjustment, the cursor will
be positioned at M10: Auto Zero Exe. (Sub-item
selection (D))
Touch the
switch to access Parameter
SET
Replacement Mode.
Touch the
INC
switch to move the cursor to
"Execution".
Touch the
switch to select the "Execution".
SET
In order to request confirmation, the entire display
flashes on and off. Touch the SET
again at this time to fix selection of the automatic
zero adjustment function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except SET
out, the parameter cannot be set.
10-2
INC
switch to
switch once
are carried
IM 01E30D01-01EN