(3) Item C (Menu C): Basic Setting items
Menu C principally contains the basic setting items for the flowtube.
Name
Item
R/W
Display unit
(BRAIN)
C00 Basic Setup
(BASIC SETUP)
C10 Tag No
(TAG NO)
C11 Flow Damping
(FLOW DAMPING)
C20 Low MF(IEL)
(LOW MF(IEL))
C21 High MF(IEL)
(HIGH MF(IEL))
C22 Low MF(IEM)
(LOW MF(IEM))
C23 High MF(IEM)
(HIGH MF(IEM))
C24 Low MF(IEH)
(LOW MF(IEH))
C25 High MF(IEH)
(HIGH MF(IEH))
C30 Nominal Size Unit
(SIZE UNIT)
C31 Nominal Size
(NOMINAL SIZE)
C40 Base Flow Unit
(FLOW UNIT)
C41 Base Time Unit
(TIME UNIT)
C42 Flow Span
(FLOW SPAN)
Data range
Units
Display unit
/BRAIN
W
ASCII 16 or ASCII 8
(*1) characters
W
1.0 to 200.0
W
0.0100 to 3.0000
W
0.0100 to 3.0000
W
0.0100 to 3.0000
W
0.0100 to 3.0000
W
0.0100 to 3.0000
W
0.0100 to 3.0000
W
mm
inch
W
0.01 to 3000.0
0.01 to 120.1
inch
W
Ml(Megaliter)
m
3
kl(Kiloliter)
l(Liter)
cm
3
m
t
kg
g
kcf
cf
mcf
Mgal (US)
kgal (US)
gal (US)
mgal (US)
kbbl (US Oil)
bbl (US Oil)
mbbl (US Oil)
ubbl (US Oil)
kbbl (US Beer)
bbl (US Beer)
mbbl (US Beer)
ubbl (US Beer)
ft
klb (US)
lb (US)
W
/s
/min
/h
/d
W
0.00001 to 32000
C40/C41
(B21 /B22)
<6. PARAMETER DESCRIPTION>
Position
Default value
of decimal
(*): Indicated item
point
s
1
5.0 s
4
1.0000 (*)
4
1.0000 (*)
4
1.0000 (*)
4
1.0000 (*)
4
1.0000 (*)
4
1.0000 (*)
mm
inch
mm
0 to 5
100 (*)
m (*)
/s (*)
0 to 4
1 m/s (*)
6-4
Description
Sets Tag number up to 16 or 8
characters.
(*1) In case of the HART
configuration tool, up to 8
characters.
Sets damping time. Linked with
B20.
Sets low-frequency meter
factor (Excitation current (Low))
Sets high-frequency meter
factor (Excitation current (Low))
Sets low-frequency meter
factor (Excitation current
(Middle))
Sets high-frequency meter
factor (Excitation current
(Middle))
Sets low-frequency meter
factor (Excitation current
(High))
Sets high-frequency meter
factor (Excitation current
(High))
Selects the nominal size units
for the flowtube.
Sets flowtube nominal size in
selected unit at C30.
Selects flow units for the flow
rate span. Linked with B21.
Selects time units for the flow
rate span. Linked with B22.
Sets flow rate span (with units
from C40 and C41). Linked
with B23.