Basic Setup; Parameters Configuration; Data Renewing - YOKOGAWA ADMAG AXR User Manual

Two-wire magnetic flowmeter integral flowmeter
Table of Contents

Advertisement

NOTE
Before updating any setting, remember to always
check the parameter content you want to change
as described in Chapter 6.
NOTE
In case of HART Configuration, the parameters
are displayed in English only.
Even if the language with the exception of
English is selected at "Language" setting display,
the parameters are displayed in English upon
HART Configuration Tool.
9.4

Basic Setup

 Confirmation and Chage of Tag and
Device Information
The tag number and device information can be
checked as follows:
• The location for the tag number and device
information
Tag
[Root Menu] → Detailed Setup
→ Basic setup → Tag
Long tag
[Root Menu] → Detailed Setup
→ Basic setup → Long tag
Descriptor
[Root Menu] → Detailed Setup
→ Device info → Field device information
→ Descriptor
Message
[Root Menu] → Detailed Setup
→ Device information
→ Field device information → Message
Date
[Root Menu] → Detailed Setup
→ Device info → Field device information
→ Date
When changing the tag number or device
information, enter the information directly within the
following limitations.
Item
Tag
Up to 8 characters or numbers*
Long tag
Up to 32 characters or numbers*
Descriptor
Up to 16 characters or numbers*
Message
Up to 32 characters or numbers*
Date
mm/dd/yyyy (DD)
yyyy/mm/dd (DTM)
- mm: month (2 digits)
- dd: days (2 digits)
- yyyy: years (4 digits)
<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
Limitations
1
2
1
1
9.5 Parameters Configuration
The parameter structure of the HART configuration
tool is hierarchical.
Read Section 9.9, Menu Tree Example for the
corresponding parameters. The menu tree shows
a cross-reference of the parameters for HART and
BRAIN.
Read Chapter 6 for the functions of parameters.
Note that some display parameters of AXR are
different from those of HART configuration tools.
9.6

Data Renewing

There are two methods to load the AXR data
from/to HART Configuration Tool --- periodic data
renewing and discretionary data renewing. Read
Section 9.9 menu tree for each corresponding
parameter.
(1) Periodic Data Renewing
The data is updated 0.5 to 2 second cycles. The
parameter of this type is marked as "D" in "Data
Renewing" of Section 9.9 menu tree.
(2) Discretionary Data Renewing
The data can be loaded from/to the AXR when
the data save is finished to the Configuration
Tool. The parameter of this type is marked as
"S" in "Data Renewing" of Section 9.9 menu
tree.
NOTE
The AXR is not able to set parameters via its
display unit during a communication of the HART
Configuration Tool.
9-3
IM 01E30D01-01EN

Advertisement

Table of Contents
loading