YOKOGAWA ADMAG AXR User Manual page 162

Two-wire magnetic flowmeter integral flowmeter
Table of Contents

Advertisement

Accuracy for Multi-drop of HART
(generally accepted values)
Size in mm (in.)
Span in m/s (ft/s)
0.3 ≤ Vs < 1
(1 ≤ Vs < 3.3)
1 ≤ Vs < 2
(3.3 ≤ Vs < 6.7)
25 to 200
(1 to 8)
2 ≤ Vs ≤ 10
(6.7 ≤ Vs ≤ 33)
Repeatability (Reference):
±0.2% of rate
(When the flow velocity is 1.5 m/s toward 2 m/s of
setting span)
Insulation Resistance:
Inspection location
Terminal
Power Supply/Digital
SUPPLY/DO
Output - Functional
Grounding
When the optional code A is selected (with the lightning protec-
tor), values are as follows.
Power Supply/Digital
SUPPLY/DO
Output - Functional
Grounding
Note: Conduct the test according to the instruction manual.
Dielectric Strength:
Inspection location
Terminal
Power Supply/Digital
SUPPLY/DO
Output - Functional
Grounding
When the optional code A is selected (with the lightning protec-
tor), values are as follows.
Power Supply/Digital
SUPPLY/DO
Output - Functional
Grounding
Note: Conduct the test according to the instruction manual.
CE Marking:
CE marking is affixed on the name plate except for
models with any of the following specifications.
• Optional Code: FF1, CF1, SF2, JF3
Safety Requirement Standards:
CAN/CSA C22.2 No. 61010-1-04
• Altitude of installation site: Max. 2000 m above sea
level
• Installation category: I
"Overvoltage category (Installation category)"
describes a number which defines a transient
overvoltage condition. It implies the regulation for
impulse withstand voltage. " I " applies to electrical
equipment which is supplied from the circuit when
appropriate transient overvoltage control means
(interfaces) are provided.
Vs: Span setting value (m/s)
Accuracy
±(0.4+0.3/Vs)% of span
± 0.5% of span
± 0.25% of span
(at indications less
than 50% of span)
± 0.5% of rate
(at indications 50% or
more of span)
Test voltage
Standard
100 MΩ
500 V DC
or more
20 MΩ
100 V DC
or more
Test
Test
Standard
voltage
time
500 V
1
25 mA
AC
min.
or less
100 V
1
6 mA
AC
min.
or less
<12. OUTLINE>
• Pollution degree: 2
"Pollution degree" describes the degree to which a sol-
id, liquid, or gas which deteriorates dielectric strength
or surface resistivity is adhering. " 2 " applies to normal
indoor atmosphere. Normally, only non-conductive
pollution occurs. Occasionally, however, temporary
conductivity caused by condensation must be expect-
ed.
• Indoor/Outdoor use
EMC Conformity Standards:
EN61326-1 Class A, Table 2 (For use in industrial
locations)
EN61326-2-3
• Performance Specification during immunity test
Flowrate output: O utput fluctuation within ±3% of
default (1m/s) span
CAUTION
This instrument is a class A product, and it is de-
signed for use in the industrial environment.
Please use this instrument in the industrial environ-
ment only.
SIL Certification:
AXR series are certified by exida in compliance with
the following standards;
IEC 61508: 2010; Part1 to Part7
Functional Safety of Electrical/electronic/programma-
ble electronic safety-related systems; SIL 2 capability
for single flowmeter use, SIL 3 capability for dual
flowmeter use.
12-7
IM 01E30D01-01EN