7 Block BASIC FUNCTION
48
7.2.3
Function group ADJUSTMENT
⇒
BASIC FUNCTION
G
BASIC FUNCTION
EPD ADJUSTMENT
Use this function to activate the EPD adjustment for an empty or full measuring tube.
(6481)
!
A detailed description of the empty pipe detection function can be found on Page 46.
Options:
OFF
FULL PIPE ADJUST
EMPTY PIPE ADJUST
Factory setting:
OFF
Procedure for EPD empty pipe / full pipe adjustment
1. Empty the piping. In case of an EPD adjustment, the wall of the measuring tube
2. Start empty pipe adjustment: Select "EMPTY PIPE ADJUST" and press F to confirm.
3. After empty pipe adjustment, fill the piping with fluid.
4. Start full pipe adjustment: Select "FULL PIPE ADJUST" and press F to confirm.
5. Having completed the adjustment, select the setting "OFF" and exit the function by
6. Now select the EMPTY PIPE DET. function (see Page 46). Switch on the empty pipe
"
The adjustment coefficients must be valid before you can activate the EPD function. If
adjustment is incorrect the following messages might appear on the display:
– FULL = EMPTY
– ADJUSTMENT NOT OK
Device Functions Proline Promag 55 FOUNDATION Fieldbus
FOUNDATION FIELDBUS GGA
⇓
PROCESS PARAMETER
Function description
→
PROCESS PARAMETER
Note!
should be wetted with fluid.
pressing F .
detection by selecting ON STANDARD and press F to confirm.
Caution!
The adjustment values for empty pipe and full pipe are identical. In such instances,
empty pipe adjustment/full pipe adjustment must be carried out again.
Adjustment is not possible because the fluid's conductivity is out of range.
⇒
GIA
CONFIGURATION
⇓
EPD PARAMETER
⇓
ADJUSTMENT
→
ADJUSTMENT
Endress+Hauser
640
642
648