14
This chapter provides instructions for replacing the following library sensors:
• Door interlock sensors (see
• Door lock solenoids (see
• I/O port open sensors (see
• I/O port solenoids (see
• Light curtain sensors (see
Important:
14.1
Replacing a Door Interlock Sensor
This section provides instructions for replacing a door interlock sensor in the XLS-832700,
XLS-820500, XLS-812300, or a MEM. Each door uses a door interlock sensor to detect if a
door is opened while the library power is off.
Procedure overview:
sensor.
Step
1
2
3
4
5
6
Table 14-1 Overview of replacing a door interlock sensor
XLS Library Technical Service Manual
Replacing Library Sensors
Section 14.2 on page
Section 14.4 on page
For clarity, all of these procedures have been
photographed with the front panel removed However,
with the exception of replacing the I/O port solenoids,
you can perform any of these procedures with the front
panel in place.
Table 14-1
Task
Shut down the library and power it off
Remove the air filter cover
Remove the door interlock sensor
Install the door interlock sensor
Replace the air filter cover
Power on the library
and Solenoids
Section
14.1)
14-4)
Section 14.3 on page
14-8)
Section 14.5 on page
provides an overview of replacing a door interlock
Refer to...
Section 6.2 on page 6-1
Section 10.1.2 on page 10-3
Section 14.1.1 on page 14-2
Section 14.1.2 on page 14-3
Section 10.2.3 on page 10-10
Section 17.3 on page 17-4
14-5)
14-10)
14-1