Purge mode
When you purge the detector, solvent flows as indicated in the following
sequence:
1.
Enters through inlet tubing port.
2.
Passes through the inlet tube of the countercurrent heat exchanger.
3.
Flows through the end-cap heat exchanger.
4.
Flows through the sample side of the flow cell.
5.
Flows out to the cross fitting, bypassing the countercurrent heat
exchanger.
6.
Flows through the reference side of the flow cell.
7.
Exits through the pressure-relief valve, and then either to waste or the
recycle valve.
October 13, 2014, 715003547 Rev. C
Features
27
Need help?
Do you have a question about the ACQUITY Refractive Index Detector and is the answer not in the manual?