Declaration-Electromagnetic Emissions
The EZ-IO® Power Driver is intended for use in the electromagnetic environment
specified below. The customer or the user of the EZ-IO® Power Driver should assure
that it is issued in such an environment.
Emission Test
Compliance
RF Emissions CISPR 11
Group 1
Class B
RF Emissions CISPR 11
Harmonic emissions IEC
Not
61000-3-2
applicable
Not
Voltage fluctuations/flicker
applicable
emissions IEC 61000-3-3
Declarations – Electromagnetic Immunity
The EZ-IO® Power Driver is intended for use in the electromagnetic environment specified
below. The customer or the user of the EZ-IO® Power Driver should assure that it is issued
in such an environment.
Immunity test
IEC 60601 test level
Compliance Level Electromagnetic Environment-Guidance
Electrostatic
discharge
+/- 8 kV Contact
+/- 8 kV Contact
(ESD)
+/- 15 kV air
+/- 15 kV air
IEC 61000-4-2
Not applicable
+/- 2 kV for power
Electrical fast
(battery
supply lines
Transient/burst
powered)
+/- 1 kV for input.
IEC 61000-4-4
Not applicable
output lines
(no I/O lines)
Not applicable
Surge
2 kV (max.)
(battery
IEC 61000-4-5
powered)
Voltage
dips, short
0% UT; 0,5 cycle
interruptions and
0% UT; 1 cycle
Not applicable
voltage variations
70% UT; 25/30 cycles
(battery
on power supply
0% UT; 250/300
powered)
input lines IEC
cycles
61000-4-11
Power frequency
(50/60Hz)
30 A/m
30 A/m
magnetic field
IEC 61000-4-8
NOTE: Ur is the a.c. mains voltage prior to application of the test level.
Compliance
The EZ-IO® Power Driver uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause
interference in nearby electronic equipment.
The EZ-IO® Power Driver is suitable for use
in all establishments, including domestic
establishments and those directly connected
to the public low voltage power supply
network that supplies buildings used for
domestic purposes.
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the EZ-IO®
Power Driver requires continued
operation during power mains
interruptions, it is recommended that
the power driver be powered from
an uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of typical location in typical
commercial or hospital enviroment.
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The EZ-IO® Power Driver is intended for use in the electromagnetic environment
specified below. The customer or the user of the EZ-IO® Power Driver should assure
that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2,7 GHz
Proximity fields
9 to 28 V/m
from RF wireless
communications
15 specific
equipment
frequencies
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters an electromagnetic site survey should be considered. If the measured
field strength in the location in which the EZ-IO® Power Driver is used exceeds the applicable RF
compliance level above, the EZ-IO® Power Driver should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the EZ-IO® Power Driver. b Over the frequency range 150 kHz to 80 MHz, field strengths
should be less than 3 V/m.
Compliance
Electromagnetic
Level
environment — guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the driver including
cables, than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter. Recommended
separation distance
d = 3,5
P
V
1
d = 3,5
P
Not applicable
E
1
(battery powered)
d = 7
P
E
10 V/m
1
where P is the maximum output
power rating of the transmitter
9 to 28 V/m
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters as determined
by an electromagnetic site
survey, a should be less than
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol: