Appendices; Fault Display And Fault Elimination; Led Flashes Red And Beeper Sounds For 3 Seconds; The Led Flashes Three Times After Pressing The Button In The Colour Of The Current Battery Capacity - Otto Bock AxonArm Hybrid 12K500 Instructions For Use Manual

Hide thumbs Also See for AxonArm Hybrid 12K500:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 21

Appendices

declaration of conformity was therefore created by the manufacturer with sole responsibility according to appendix
VII of the guidelines.
12 Appendices

12.1 Fault Display and Fault Elimination

12.1.1 LED flashes red and beeper sounds for 3 seconds

Probable cause:
Error in the Axon-Bus prosthetic system.
Remedy:
1. Switch the Axon-Bus prosthetic system off and back on.
2. Check the plug connections in the socket.
3. Connect the charger. If the charger displays the service symbol (yellow spanner) or the
12.1.2 The LED flashes three times after pressing the button in the colour of the current battery
capacity
Probable cause:
The connection from the AxonEnergy Integral to the Axon-Master has been interrupted.
Remedy:
1. Check the connection (e.g. plug connections, flex cable, EasyPlug etc.).
2. If no fault can be found, send the Axon-Bus prosthetic system to Ottobock Myo-Service.

12.1.3 LED alternately flashes yellow/red and beeper sounds for 3 seconds

Probable cause:
The Axon-Bus prosthetic system is overheated.
Remedy:
1. Allow the Axon-Bus prosthetic system to cool for a few minutes.
2. Do not use the Axon-Bus prosthetic system at high ambient temperatures.

12.1.4 The movements of the Axon-Bus gripping component are becoming increasingly slower

The patient is alerted to the diminishing battery charge level as the Axon-Bus gripping component increasingly
slows down and the gripping force decreases. Finally, when there is very little battery capacity remaining, the
Axon-Bus prosthetic system switches off automatically to protect the battery against harmful deep discharge.

12.2 Symbols Used

12.3 Technical Data

Description
Operating temperature
Storage and transport in original
packaging
36 | Ottobock
system error symbol (red spanner) send the Axon-Bus prosthetic system to Ottobock Myo-
Service.
Declaration of conformity according to the European Directive for Medical Devices
93/42/EEC.
This product may not be disposed of with regular domestic waste in all jurisdictions.
Disposal that does not comply with the regulations of your country may have a detri­
mental impact on health and the environment. Please observe the instructions
provided by the responsible authorities in your country regarding return and collection
processes.
Serial number of device
Legal manufacturer
12K500=* AxonArm Hybrid
0 °C to 60 °C/32 °F to 140 °F
-20 °C to 40 °C/-4 °F to 104 °F
12K501=* AxonArm Ergo
0 °C to 60 °C/32 °F to 140 °F
-20 °C to 40 °C/-4 °F to 104 °F
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*

Advertisement

Table of Contents
loading

This manual is also suitable for:

Axonarm ergo 12k501

Table of Contents