Otto Bock DynamicArm 12K100N Instructions For Use Manual

Otto Bock DynamicArm 12K100N Instructions For Use Manual

Hide thumbs Also See for DynamicArm 12K100N:

Advertisement

Quick Links

DynamicArm 12K100N
Instructions for use (qualified personnel) .................................................................
3

Advertisement

Table of Contents
loading

Summary of Contents for Otto Bock DynamicArm 12K100N

  • Page 1 DynamicArm 12K100N Instructions for use (qualified personnel) ..............
  • Page 2 Εκτός από τις τυπωμένες οδηγίες χρήσης, συμπεριλαμβάνονται και άλλες γλώσσες σε CD (βλ. πίσω εξώφυλλο). Μπορείτε να παραγγείλετε δωρεάν τις έντυπες οδηγίες χρήσης στη γλώσσα της χώρας σας από τη διεύθυνση που αναφέρεται παρακάτω. Ottobock Healthcare Products GmbH Brehmstraße 16 | 1110 Wien | Austria Service-admin.vienna@ottobock.com | Fax (+43-1) 526 79 85 DynamicArm 12K100N...
  • Page 3: Product Description

    The free swing of the forearm corresponds to the natural movement behaviour of the upper limb while walking. After extension of the forearm, the vario-gear automatically decouples the forearm from the drive unit so it can DynamicArm 12K100N...
  • Page 4 With the product switched off or when the battery is drained, the elbow can be manually unlocked and locked again with a slight pull on the unlock cable. In this way, the forearm can be brought into the desired position. Mechanical release can also take place under load. DynamicArm 12K100N...
  • Page 5 The fitting of a patient with the product may only be carried out by O&P professionals who have been authorised with the corresponding Ottobock training. 4 Safety 4.1 Explanation of warning symbols Warning regarding possible serious risks of accident or injury. WARNING Warning regarding possible risks of accident or injury. CAUTION Warning regarding possible technical damage. NOTICE DynamicArm 12K100N...
  • Page 6: Structure Of The Safety Instructions

    ► If the product and its components have been subjected to extreme loads (e.g. due to a fall, etc.), then the product must be inspected for damage immediately. If necessary, forward the product to an authorised Ottobock Service Centre. DynamicArm 12K100N...
  • Page 7 CAUTION Mechanical stress on the prosthesis Injury due to faulty control or malfunction of the prosthesis. ► Do not subject the prosthesis to mechanical vibrations or impacts. ► Check the prosthesis for visible damage before each use. DynamicArm 12K100N...
  • Page 8 ► Ensure that fingers and other body parts are not in this area when bending the joint. NOTICE Coating, gluing or painting the prosthesis Damage or fracture due to chemical processes. ► The prosthesis must not be coated, glued or painted. DynamicArm 12K100N...
  • Page 9: Scope Of Delivery And Accessories

    If the cables were pulled out anyway, they have to be pushed back into the forearm and secured with the foam disc before performing further work. The cables take up more space in the forearm as a result. DynamicArm 12K100N...
  • Page 10 2) Cut the forearm using a vibrating saw. 3) Sand all sides of the cut edge at a right angle. Smooth inner and outer sanded edges. Round the inside edge slightly. 4) Using the hook, remove the foam cover. DynamicArm 12K100N...
  • Page 11 2) Mark this measurement on the inside of the fore­ arm. 3) Mix 636K18=1 Orthocryl sealing resin compact adhesive and 2% 617H14 hardener paste. 4) Cover the outside of the forearm with 627B4 poly­ ethylene adhesive tape to avoid soiling it while glu­ ing. DynamicArm 12K100N...
  • Page 12 For forearm lengths from 225 mm to 187 mm without 10S17 electric wrist rotator 1) Pull the 10S17 electric wrist rotator dummy back and out of the lamination ring. The lamination ring cover has to remain in the lamination ring. DynamicArm 12K100N...
  • Page 13 3) Or, pull out the lamination ring cover on its own. 4) Remove the cable protector cap using the hook. 5) The cables are rolled up inside the cable protector cap. When removing the cable protector cap, the cables are pulled out of the forearm. DynamicArm 12K100N...
  • Page 14 2) Lubricate the groove, inner edges and thread of the lamination ring with 633F30 special grease. 6.5 Plugging in the contact 1) Apply a generous amount of 633F11 silicone grease to the connector contacts of the 13E219 control electronics. DynamicArm 12K100N...
  • Page 15: Assembling The Housing

    2) The circuit board and the edge of the lamination ring must be parallel to each other. 3) The electrode housing has engaged properly. The circuit board and the edge of the lamination ring are parallel to each other. DynamicArm 12K100N...
  • Page 16 3) Set the AFB dial to minimum compensation force. 4) Extend the elbow and lock it with the pull cable. Switch off the product. 5) Remove the Allen head screw. DynamicArm 12K100N...
  • Page 17 For safety reasons, it must not be removed! 3) Rotate the socket in and out up to the stop. Ensure that the pull cable does not tighten. Complete the following additional steps to finish the switch cable: DynamicArm 12K100N...
  • Page 18 ► The upper arm rotation joint has bilateral stops (± 80 degrees) to block extreme movements. The fric­ tion of the humeral rotation feature at the upper arm connection is easily adjusted by turning the external adjustment screw. DynamicArm 12K100N...
  • Page 19 6.12 Wiring without electric wrist rotator 1) One of the two three-pole connectors is marked with a coloured dot (arrow). The two receptacles must not be interchanged! INFORMATION: The two-pole motor cable must not be attached to the coaxial plug. DynamicArm 12K100N...
  • Page 20: Charging The Battery

    8. Disconnect the power cord of the battery charger from the wall socket. 7.1.2 Charging times Charging time Capacity Operating time 3 hours 100% approx. 18 hours 1.5 hours approx. 14 hours DynamicArm 12K100N...
  • Page 21 The elbow joint cannot be adjusted if the calibration process has not been completed yet. 1) Mount the product on the socket. 2) Switch on the product. 3) Start the “646C42=V1.6 ElbowSoft” software. For the subsequent steps, see the corresponding instructions for use. DynamicArm 12K100N...
  • Page 22: Switching Modes

    Switching using long co- Switching to the hand Switching to the elbow Switching to the electric contraction joint wrist rotator Switching using short Switching to the electric Switching to the hand Switching to the elbow co-contraction wrist rotator joint Switching back DynamicArm 12K100N...
  • Page 23 Automatically switching back to the hand component takes place five seconds after relaxing the muscles. DynamicArm Electric wrist rotator System Electric Hand and System Electric Gre­ ifer Control Linear control element Two electrodes Indication For patients with two strong muscle signals Control type Position control Digital DynamicArm 12K100N...
  • Page 24 Linear control element Two electrodes Indication • For patients with two muscle signals of any strength • Possibility for simultaneous control of elbow joint and terminal device Control type Position control Proportional Switching Switching to the elbow joint DynamicArm 12K100N...
  • Page 25: Cleaning And Care

    Regular service inspections (maintenance) are mandatory in the interest of patient safety and in order to maintain operating reliability and protect the warranty. These service inspections include inspecting the sensors and replacing wear and tear parts. See the service pass for further information about the prescribed service intervals. DynamicArm 12K100N...
  • Page 26: Local Legal Information

    Caution: Exposure to Radio Frequency Radiation. This device must not be co-located or operating in conjunction with any other antenna or transmitter. Responsible party: Otto Bock Health Care, LP 3820 West Great Lakes Drive Salt Lake City, Utah 84120-7205 USA...
  • Page 27: Technical Data

    RF field in excess of Health Canada limits for the general population; consult Safety Code 6, obtainable from Health Canada’s website http://www.hc-sc.gc.ca/rpb. Responsible party: Otto Bock Healthcare Canada Ltd. 5470 Harvester Road L7L 5N5 Burlington, Ontario Canada Phone + 1-800-665-3327 Caution: Federal law (USA) restricts this device to sale by or on the order of a practitioner licensed by law of the State in which he/she practices to use or order the use of the device.
  • Page 28: Symbols Used

    Contact 1 x long Severe error Ottobock Service if the (e.g. temperature of the lift­ error been ing motor too high) resolved. Malfunction (e.g. incompatible com­ ponent connected) DynamicArm 12K100N...
  • Page 29 6 x low Battery charger No function No function No function was connected during opera­ tion. Low, high Prosthetic com­ Function of the prosthetic component in question is ponent is being not possible. controlled. DynamicArm 12K100N...
  • Page 30 DynamicArm 12K100N...
  • Page 31 DynamicArm 12K100N...
  • Page 32 Otto Bock Healthcare Products GmbH Brehmstraße 16 · 1110 Wien · Austria T +43-1 523 37 86 · F +43-1 523 22 64 info.austria@ottobock.com · www.ottobock.com...