INFORMATION
Switching off the Axon-Bus prosthetic system during extended passive pauses (e.g. air or rail travel, visit to a
theatre or cinema, etc.) will make the battery last longer before requiring a recharge. Only the entire Axon-Bus
prosthetic system with all connected components can be switched off. Individual Axon-Bus components cannot
be switched off separately.

2.8 Charging Receptacle

The charging receptacle (Fig. 1, item 2) has the following functions when the AxonArm Ergo is connected with the
AxonMaster:
Contacts for charging the battery
LED display to indicate the current battery charge level
LED display for providing feedback on operating states
Button to turn the Axon-Bus prosthethic system on and off, display the battery charge level and open the Axon-
Bus gripping component in an emergency
Beeper for providing feedback on operating states
Button for activating the Bluetooth function

2.9 Rechargeable Battery

The rechargeable battery (Fig. 1, item 1) consists of Li-ion cells with integrated electronics. This protects the bat­
tery against short circuits, overvoltage, undervoltage and charging outside the allowable temperature range.

3 Application

3.1 Indications for Use

The AxonArm Hybrid 12K500 and the AxonArm Ergo 12K501 may only be used exclusively for exoprosthetic fit­
tings of the upper limbs.

3.2 Orthotist / Prosthetist Qualifications

Fitting a patient with the AxonArm may only be carried out by prosthetists who have been authorised by Ottobock
after completion of a corresponding training course.

3.3 Conditions of Use

The AxonArm was developed for everyday use and must not be used for unusual activities. These unusual activities
include, for example, sports with excessive strain and/or shocks to the wrist unit (pushups, downhill, mountain bik­
ing) or extreme sports (free climbing, paragliding, etc.).
The applicable environmental conditions are described in the technical data.
The AxonArm is intended for use by one user only. Use of the product by another person is not approved by the
manufacturer.

3.4 Application Area

The AxonArm is intended solely for use with components of the Axon-Bus prosthethic system. It can be used for
unilateral or bilateral amputees from a transhumeral amputation level to shoulder disarticulation.

4 Safety

4.1 Explanation of Warning Symbols

Warning regarding possible serious risks of accident or injury.
WARNING
Warning regarding possible risks of accident or injury.
CAUTION
Warnings regarding possible technical damage.
NOTICE
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*
Application
Ottobock | 25