Installing The Column - Waters ACQUITY UPLC Operator's, Overview And Maintenance Manual

Table of Contents

Advertisement

2.
Remove the column in-line filter unit from its packaging, and then push
the outlet end of the active preheater or column stabilizer tubing
assembly into the in-line filter inlet until it stops.
Installing the in-line filter:
Column
3.
Holding the reusable fitting in place, rotate the in-line filter unit onto
the fitting until it is snug.
Caution:
tighten the tubing fittings.
4.
Finger tighten the fitting until it is snug, and then tighten it up to an
additional 1/6-turn.

Installing the column

Warning:
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column compartment
internal temperature to ensure all components are cool.
Caution:
chemical-resistant, powder-free gloves when installing the column.
The procedure for installing the column varies depending on the column
heater model.
To avoid damaging the tubing, use only your fingers to
To prevent burn injuries, set the column temperature to Off,
To prevent contaminating system components, wear clean,
Flow
In-line filter unit
Inlet
Preparing the column compartments
From column stabilizer
tubing assembly
TP02873
Reusable fitting