Column stabilizer assembly (two styles):
For 150-mm column
8.
Carefully route the column-stabilizer-inlet tubing through the groove on
the side of the column compartment.
Caution:
9.
Insert the Phillips screws into the threaded holes, and tighten them
gently using the Phillips screwdriver.
10. Route the column-stabilizer-inlet tubing through the cutout in the top of
the sample manager door to the injector.
Do not route the tubing behind the sample manager door
Requirement:
hinge.
Reusable fittings
TP03123
To avoid stripping threads, do not overtighten screws.
Maintaining the column compartments
For 50-mm or 100-mm column
Inlet tubing to injector
135