Waters ACQUITY UPLC Operator's, Overview And Maintenance Manual page 51

Table of Contents

Advertisement

Two-column configuration plumbing connection details: (Continued)
From
To
CM-A inlet
Column 1 inlet
valve, port 3
CM-A inlet
Column 2 inlet
valve, port 4
Column 1 outlet CM-A outlet
valve, port 3
Column 2 outlet CM-A outlet
valve, port 4
Sample
CM-A inlet
manager
valve, center
injector valve,
port
port 6
Detector inlet
CM-A outlet
valve, center
port
a. Be sure to connect the 430002727 waste tubing to port W (at the 3 o'clock position) of the inlet
valve. Do not connect to the center port.
Length
(inches)
H-Class
12.5
205000730,
APH
12.5
205000730,
APH
19
430002683, SST
tubing,
.004-inch ID
(reorder part
number
700005478)
19
430002683, SST
tubing,
.004-inch ID
(reorder part
number
700005478)
14.5
430002582, SST
tubing,
.004-inch ID
(reorder part
number
700005476)
14.5
430002681, SST
tubing,
.004-inch ID
(reorder part
number
700005477)
Preparing the column compartments
Part number
I-Class
205000894,
APH
205000894,
APH
430003151, SST
tubing,
.003-inch ID
(reorder part
number
700008868)
430003151, SST
tubing,
.003-inch ID
(reorder part
number
700008868)
430003154, SST
tubing,
.003-inch ID
(reorder part
number
700008869)
430003158, SST
tubing,
.003-inch ID
(reorder part
number
700008870)
47

Advertisement