7.
Plug the leak sensor connector into the receptacle on the front of the
instrument.
8.
Power-on the column manager.
9.
In the ACQUITY UPLC Console, select Column Manager from the
system tree.
10. In the column manager information window, click Control > Reset CM,
to reset the column manager.
Replacing the column manager switching-valve cartridge
Before replacing the switching-valve cartridge, ensure you have the correct
part for your system type (ACQUITY regular or Bio).
Warning:
including inhalation, observe Good Laboratory Practice when you
handle them. See the Material Safety Data Sheets for the solvents you
use.
Caution:
chemical-resistant, powder-free gloves when replacing the column
manager switching-valve cartridge.
122
Front of instrument
To avoid the harmful effects of personal contact with solvents,
To prevent contaminating system components, wear clean,
Leak sensor
connector receptacle