b.
Using two TORX screws, attach the height adjustment bracket to
the MS-bracket.
c.
Using two TORX screws, attach the hinge assembly on the column
heater housing to the height adjustment bracket.
d.
Swing the column heater housing, aligning it with the front of the
mass spectrometer, and ensure that the column heater is at the
correct height for the mass spectrometer inlet.
e.
If the column heater housing is not at the correct height, adjust the
height by removing the hinge from the height adjustment bracket
and then reattaching it using different holes.
19. Attach the MS drip tray to the column heater housing using the Phillips
screws.
Leak sensor
cable slot
Drip tray
Installing the leak sensor is optional.
Tip:
TORX screws
MS-bracket
Locations of Phillips screws
Preparing the column compartments
Height
adjustment
bracket
27