14. Align the leak sensor's T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
T-bar
Leak sensor
installed in
reservoir
15. If you detached the connector from the front of the instrument, reattach
it, and power-on the sample manager.
16. In the ACQUITY UPLC Console, select Sample Manager FTN from the
system tree.
17. In the sample manager information window, click Control > Reset SM,
to reset the sample manager.
To resolve a 30-cm column heater leak sensor error:
1.
In the ACQUITY UPLC Console's Leak Sensors dialog box, determine
which of the sample manager's two leak sensors detected a leak.
2.
If the message reads "Leak Detected (Column)", open the 30-cm column
heater's door.
3.
Locate the source of the leak, and make the repairs necessary to stop it.
104
TP02892
Slot in leak
sensor reservoir