Replacing the column heater's leak sensor

Perform this procedure if you must replace the leak sensor.
Required materials
Gloves: clean, powder-free, chemical-resistant
New leak sensor
Required tools
Phillips screwdriver
T20 TORX driver
To replace the column heater's leak sensor:
1.
Power-off the sample manager.
2.
Disconnect the column inlet, active preheater, or column stabilizer
tubing from injector port 6 of the sample manager.
114
To avoid personal contamination from contact with
Warning:
biologically hazardous or toxic materials, always wear clean,
chemical-resistant, powder-free gloves when performing this
procedure.
Location of column, active
preheater, or column-stabilizer-
inlet tubing