Installing Or Replacing A Column Plate In The Column Compartment - Waters ACQUITY CM-30S Overview And Maintenance Manual

30-cm single zone column manager
Table of Contents

Advertisement

3.
Insert the stainless-steel tubing, with ferrule, into the column.
4.
Seat the ferrule by finger tightening the dual-threaded fitting.
Requirement:
seating the ferrule.
5.
Complete the connection by finger tightening the cap nut onto the
dual-threaded fitting.
Installing or replacing a column plate in the column
compartment
The illustrations in the following procedure depict a column installed in
Note:
position 3 on a column plate in a UPC
replacing a column plate in an APC system, the column bank will be attached
at either fitting position 1a or 2a.
Warning:
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column-compartment
internal temperature to ensure all components are cool.
Notice:
chemical-resistant, powder-free gloves when installing a column plate.
Required materials
Column support clips
Gloves: clean, powder-free, chemical-resistant
Reusable fitting

Installing or replacing a column plate in the column compartment

Do not rotate the stainless-steel silver cap nut when
To prevent burn injuries, set the column temperature to Off,
To prevent contaminating system components, wear clean,
2
system. If you are installing or
October 7, 2014, 715004178 Rev. B
45