6.
Roll up a nonabrasive, lint-free wipe, and use it to absorb the liquid from
the leak sensor reservoir and its surrounding area.
Leak sensor
reservoir
7.
With a cotton swab, absorb any remaining liquid from the corners of the
leak sensor reservoir and its surrounding area.
Leak sensor
reservoir
106
Rolled up
lint-free wipe
Cotton swab