Repositioning instruments in the system stack
Requirement:
column module, the CM-A or the CM-Aux, is fitted with the drip tray equipped
with a leak detector (see
To reposition the instruments in the system stack:
1.
Power-off the column manager (CM-A) and detector.
2.
Disconnect the power and Ethernet cables from the rear panel of the
CM-A and detector.
3.
Remove all solvent bottles from the bottle tray.
4.
Remove the solvent tray from the system stack.
5.
Disconnect the waste tubing from the detector and the inlet tubing from
the CM-A outlet.
6.
Remove the detector from the system stack.
7.
Remove the CM-A waste tubing from the solvent manager drip tray.
8.
Remove the sample manager outlet tubing from port 6 of the injector.
9.
Arrange the instrument modules in the new configuration.
10. Place the solvent tray in the appropriate location in the system stack.
11. Place the solvent bottles in the solvent tray.
12. Connect the power cables to the upper CM-Aux, CM-A, lower CM-Aux (if
present), and detector.
13. Connect the external control cables to the rear of the CM-A as follows:
Upper auxiliary port to upper CM-Aux
Lower auxiliary port to lower CM-Aux (if present)
14. Connect the Ethernet cables at the rear panels of the CM-A and
detector.
58
When repositioning the instruments, ensure the bottommost
page
To avoid injuries arising from contact with spilled
Warning:
solvent, remove the solvent bottles from the solvent tray before
removing the solvent tray.
59).