c.
Tighten the cable clamp nut with the 4-mm wrench.
25-pin cable routed
through slot
4-mm cable clamp
nut on stud
Cable clamp
22. Install the column heater in its housing:
a.
Insert the tab on the left-hand side of the column heater into the slot
on the left-hand side of the housing.
b.
Hold the left-hand side of the column heater into the slot, and guide
the eCord tab, on the right-hand side, into the cutout in the column
heater housing.
c.
Install the TORX screw in the tab and tighten it.
Preparing the column compartments
Tab
Cutout