7.
Insert the fuse holder into the receptacle, and gently push until it locks
into position.
A clicking sound indicates the fuse holder is locked into position.
Tip:
8.
Connect the power cord to the rear panel power entry module.
Cleaning the instrument's exterior (all models)
Warning:
the column manager to a safe temperature (10 to 30 °C) and allow the
unit to attain this set-point temperature before handling components
within the column compartment. When the column manager is set at a
high temperature, all interior surfaces and components are hot.
Clean surfaces of the column heater, column manager, auxiliary column
manager, or column heater/cooler with only a soft, lint-free paper or cloth
dampened with water.
Observe these requirements when cleaning device surfaces:
Always ensure the electrical power to the device is interrupted.
Always use eye and hand protection during the cleaning process.
Apply the water to a clean cloth only, and then wipe the device.
Never spray or apply the water directly onto any device surface.
140
To prevent burn injuries, adjust the temperature set-point of