Waters ACQUITY UPLC Operator's, Overview And Maintenance Manual page 92

Table of Contents

Advertisement

Required materials
Column in-line filter
Gloves: clean, powder-free, chemical-resistant
Reusable fitting
Required tools
5/16-inch open-end wrench
For instruction on how to replace the filter in the column heater's in-line filter
unit, see
page
Caution:
powder-free, non-latex gloves when connecting tubing and tightening
fittings.
To install the column in-line filter:
1.
Remove the O-ring retainer from the outlet end of the active preheater
or column stabilizer tubing.
Before connecting, make sure that the ferrule is approximately
Tip:
3 mm from the end of the outlet tubing and that the collet is touching
the back of the ferrule.
Ferrule
3 mm
Older, reusable fittings do not have holes in their sides.
88
136.
To prevent contaminating system components, wear
Reusable fitting
Collet
Outlet tubing