Align the leak sensor's T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
T-bar
Leak sensor
installed in
reservoir
9.
In the ACQUITY UPLC Console, select Sample Manager FTN from the
system tree.
10. In the sample manager information window, click Control > Reset SM,
to reset the sample manager.
To resolve a column manager leak sensor error:
1.
Power-off the column manager.
2.
Open the column manager door.
3.
Locate the source of the leak, and make the repairs necessary to stop it.
Liquid from a detector leak can accumulate in the column manager
Tip:
leak sensor reservoir and cause the column manager leak sensor alarm
to sound. Always inspect the detector when attempting to locate the
source of a leak.
TP02892
Slot in leak
sensor reservoir