Signal Mode; Detection Software Upgrade - Sirona PerioScan Operating Instructions Manual

Hide thumbs Also See for PerioScan:
Table of Contents

Advertisement

Available languages

Available languages

6 Detection mode (calculus detection)
6.4

Signal mode

100
100
1%
25%
6.5

Detection software upgrade

The existing detection software must be upgraded soon as new detection tips become available.
22
PERIO 1
PERIO 2
PERIO 3
50%
75%
100%
25
A
The user has the option of activating a signal tone if the
detection function is activated. This feature can be used
to detect calculus as well as to indicate that the distal
end of the tip is not located on the root surface.
The signal mode can be activated by simultaneously
pressing the "Small rinsing media tank" and "PERIO 1"
keys. As soon as the signal mode is activated, the tooth
displayed on the screen starts flashing. Here, it is possi-
ble to adjust the volume of the signal with the 50 % and
100 % keys. Furthermore, the signal tone can now be
assigned to the detection of calculus with the "PERIO 1"
key or to noncontact of the tip with the root surface with
the "PERIO 2" key. The user can see which function he
has activated based on the readings that appear on the
display following his selection (doctoral cap = signal set
to noncontact of tip with root surface).
The signal tone can be deactivated (detection of calcu-
lus or noncontact of tip on root surface) by pressing the
activated key once again. You can quit the signal mode
by simultaneously pressing the "Small rinsing media
tank" and "PERIO 1" keys once again.
Proceed as follows:
Switch off the unit.
Insert memory card A in the card reader (25) as
shown until it snaps into place.
Switch the unit on.
The display reads "100".
A countdown begins and decrements to "1".
Then a welcome screen displaying the updated soft-
ware version automatically appears.
Briefly tap and remove memory card A.
The unit is ready for operation.
ATTENTION
Do not switch the unit off during the upgrade process!
Do not remove the memory card during the upgrade pro-
cess!
Sirona Dental Systems GmbH
Operating Instructions PerioScan
61 29 626 D 3496
D 3496.201.02.05.02

Advertisement

Table of Contents
loading

Table of Contents