[SOURce:]MEMory:LIST:DATA?

Syntax
[SOURce:]MEMory:LIST:DATA? (@<ch>)
This command reads the data result after executing the commands in the channel
active memory list through a remote or hardware trigger action.
Parameter
Item
<ch>
Remarks
This command is only applicable for the U2723A USB Modular SMU.
Returned Query Format
<NR3>{,<NR3>}
The reading is in the form of +9.99999999E+10 when output is set to OFF or no
measurement is made during the memory list's commands execution.
Array values responses are separated by commas.
Example
MEM:LIST:DATA? (@1)
See Also
[SOURce:]CURRent:RANGe
[SOURce:]MEMory:LIST:CLEar
[SOURce:]MEMory:LIST:POINts?
[SOURce:]MEMory:LIST:LOAD
[SOURce:]MEMory:LIST:READ?
[SOURce:]MEMory:LIST:STORe
Keysight U2722A/U2723A Programmer's Reference
Type
NR1
Range of values
1 through 3
SOURce Subsystem
Defaul t value
Required parameter
9