11
SYSTem Subsystem
SYSTem:CDEScription?
Syntax
SYSTem:CDEScription?
This query is used when the SMU device is slotted into modular instrument
chassis. It identifies which slot is the module plugged into.
Remarks
– It is only applicable when the U2722A/U2723A is used in the modular
– If the U2722A/U2723A is not plugged into the chassis, then the returned query
– Please refer to the U2781A Modular Instrument Chassis User's Guide for the
Returned Query Format
<NR1>, <NR1>
Examples
The following shows the query when the U2722A/U2723A is slotted into slot 4 of
chassis #1.
SYST:CDES?
Typical Response: +4, +1
Below is the query when the U2722A/U2723A is plugged into slot 2 of chassis #2.
SYST:CDES?
Typical Response: +2, +2
104
instrument chassis.
value would be "+7, +0", where "+7" is an invalid slot number and "+0" is the
chassis number.
details.
Keysight U2722A/U2723A Programmer's Reference