[Source:]Memory:configure:points - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

[SOURce:]MEMory:CONFigure:POINts

Syntax
[SOURce:]MEMory:CONFigure:POINts <start>,<end>,<loop>, (@<ch>)
This command sets the start point, end point, and loop count for the channel
active memory list.

[SOURce:]MEMory:CONFigure:POINts?

Queries the start point, end point, and loop count value for the channel active
memory list.
Parameter
Item
Type
NR1
<ch>
NR1
<start>
NR1
<end>
NR1
<loop>
Remarks
This command is only applicable for the U2723A USB Modular SMU. If the set
value is larger than the number of commands in the memory list, or the end point
is larger than the start point,the input is invalid and error –222 is displayed.
Returned Query Format
<NR1>,<NR1>,<NR1>
Example
MEM:CONF:POIN 1,100,1(@1)
MEM:CONF:POIN?(@1)
See Also
CONFigure:SSI
Keysight U2722A/U2723A Programmer's Reference
Range of values
1 through 3
1 to 200
1 to 200
1 to 1000
SOURce Subsystem
Defaul t value
Required parameter
Required parameter
Required parameter
Required parameter
9