Error Messages; Error List; Table 14-1 Device-Dependent Errors - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

14

Error Messages

Error Messages
Error messages are created once a command error or an erroneous condition has
been detected.
– Errors are retrieved in first-in, first-out (FIFO) order.
– Errors are cleared as you read them.
– No additional errors will be stored until you remove the errors from the queue.

Error List

The table below contains the device-dependent errors.
Table 14-1
Device-dependent errors
Error
Device-dependent errors (these errors set the Standard Event Status register bit #3)
0
No error
This is the response to the
+101
Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
+103
Calibration vol tage or current range is incorrect
Calibration voltage and current ranges are set incorrectly. Calibration cannot be executed.
+104
Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
+120
Configuration Mul tiple slave not allowed
Multiple slaves are not allowed in the configuration.
+121
Configuration master not allowed
Status master is not allowed in the configuration.
+246
Master Controller Busy
The channel is busy. All input commands will be ignored.
130
If no errors occur when you read the error queue, the instrument will respond
with 0, "No error".
SYSTem:ERRor?
will read and clear one error from the queue.
SYSTem:ERRor?
query when there are no errors.
Keysight U2722A/U2723A Programmer's Reference