Sense:current[:Dc]:Nplcycles - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

SENSe:CURRent[:DC]:NPLCycles

Syntax
SENSe:CURRent[:DC]:NPLCycles <cycles>, (@<ch>)
This command sets the number of power line cycles taken for the current
measurement. The value is coupled to APERture by the equation:
SENSe:CURRent[:DC]:NPLCycles? (@<ch>)
Query the number of power line cycles taken for the current measurement in the
defined channels.
Parameters
Item
<ch>
<cycles>
Returned Query Format
<NR1>
Examples
SENS:CURR:NPLC 10, (@1)
SENS:CURR:NPLC? (@1)
See Also
SENSe:CURRent[:DC]:APERture?
SENSe:VOLTage[:DC]:NPLCycles
Keysight U2722A/U2723A Programmer's Reference
APERture = NPLCycles/selected line frequency.
Type
NR1
NR1
Range of values
1 through 3
0 to 255
SENSe Subsystem
Defaul t value
Required parameter
Required parameter
8