[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]

Syntax
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <current>,
(@<ch>)
This command sets the current trigger level of the specified output channel. Units
are in amperes. The triggered level is a stored value that is transferred to the
output when an output step is triggered.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? (@<ch>)
Queries the stored value of triggered current.
Parameters
Item
Type
NR1
<ch>
NRf
<current>
Returned Query Format
<NR3>
Examples
The following command sets the current trigger level to 0.02 A on channel 1.
CURR:TRIG 0.02, (@1)
The following query returns the current trigger level on channel 1.
CURR:TRIG? (@1)
Typical Response: +2.000000E-02
See Also
INITiate[:IMMediate]:TRANsient
[SOURce:]CURRent:RANGe
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Keysight U2722A/U2723A Programmer's Reference
Range of values
1 through 3
The maximum value is dependent on the current range set
SOURce Subsystem
Defaul t value
Required parameter