10
STATus Subsystem

STATus:OPERation:NTRansition

Syntax
STATus:OPERation:NTRansition <value>
This register serves as a polarity filter between the Operation Condition and
Operation Event registers. When a bit in the Operation NTR register is set to 1,
then a 1-to-0 transition of the corresponding bit in the Operation Condition
register causes that bit in the Operation Event register to be set.

STATus:OPERation:NTRansition?

These command reads the value of the Operation NTR (Negative-Transition)
register.
For more information on the SCPI Status System, refer to
NOTE
system diagram"
Parameters
Item
Type
NR1
<value>
For example, to enable bit 5 (decimal value = 32) and bit 6 (decimal value = 64),
the corresponding decimal value would be 96 (32 + 64).
88
on page 21.
Range of values
A decimal value which corresponds to the binary-weighted
sum of the bits in the register (refer to the table below).
Keysight U2722A/U2723A Programmer's Reference
Chapter
2,
"Status
Defaul t value
Preset = 0