[Source:]Voltage:limit - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

[SOURce:]VOLTage:LIMit

Syntax
[SOURce:]VOLTage:LIMit <value>, (@<ch>)
This command sets the maximum bounds on the output voltage value. Output
voltage level will be clamped to the limit value if the voltage level has exceeded
the bounds set.
[SOURce:]VOLTage:LIMit? (@<ch>)
This will query the set voltage limit.
Parameters
Item
Type
NR1
<ch>
NRf
<value>
Returned Query Format
<NR3>
Examples
The following command sets the voltage limit to 0.5 V on channel 1.
VOLT:LIM 0.5, (@1)
The following query returns the programmed voltage limit on channel 1.
VOLT:LIM? (@1)
Typical Response: +5.000000E-01
See Also
[SOURce:]CURRent:LIMit
Keysight U2722A/U2723A Programmer's Reference
Range of values
1 through 3
The maximum value is dependent on the
voltage range set
SOURce Subsystem
Defaul t value
Required parameter
Required parameter
9
79

Hide quick links: