Ese/*Ese - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

13
IEEE-488.2 Common Commands

*ESE/*ESE?

Syntax
*ESE <enable_value>
This command enables bits in the enable register for the Standard Event register
group. The selected bits are then reported to bit 5 of the Status Byte register.
*ESE?
This will query the Standard Event register group.
Parameter
Item
<enable_value>
Remarks
– The following table lists the bit definitions for the Standard Event register.
Bit number
0 Operation Complete
1 Not Used
2 Query Error
3 Device-Dependent Error
4 Execution Error
5 Command Error
6 Not Used
7 Power On
114
Type
Range of values
NRf
A decimal value that corresponds to the binary-weighted
sum of the bits in the register. Refer to the table below.
Decimal value
Definition
1
All commands prior to and including *OPC have been executed.
2
Always zero.
4
The instrument tried to read the output buffer but it was empty. Or, a new
command line was received before a previous query has been read. Or, both
the input and output buffers are full.
8
A self-test or calibration error occurred. For a complete listing of the error
messages, refer to
16
An execution error occurred (an error in the –200 range has been generated.
32
A command syntax error occurred (an error in the –100 range has been
generated).
64
Always zero.
128
Power has been turned off and on since the last time the event register was
read or cleared.
Chapter
14,
"Error Messages"
Keysight U2722A/U2723A Programmer's Reference
Defaul t value
This is a required
parameter.
on page 130.

Hide quick links:

Advertisement

Table of Contents
loading