Remarks
– The following table lists the bit definitions for the Operation Status registers.
Bit number
Decimal value
0 to 1 Not Used
0
2 DTG1-tran
4
3 DTG2-tran
8
4 DTG3-tran
16
5 WTG1-tran
32
6 WTG2-tran
64
7 WTG3-tran
128
8 to 15 Not Used
0
– The
Returned Query Format
<NR1>
The returned value is the binary-weighted sum of all bits set in the register. For
example, if bit 5 (decimal value = 32) and bit 6 (decimal value = 64) are set, the
query command will return "+96".
Examples
The following command enables bits 5 (decimal value = 32) and 6 (decimal
value = 64) in the enable register.
STAT:OPER:ENAB 96
The following query returns the bits enabled in the register.
STAT:OPER:ENAB?
Typical Response: +96
See Also
STATus:OPERation[:EVENt]?
STATus:PRESet
Keysight U2722A/U2723A Programmer's Reference
Definition
0 is returned.
The transient system has received a trigger signal and operation is running in channel 1.
The transient system has received a trigger signal and operation is running in channel 2.
The transient system has received a trigger signal and operation is running in channel 3.
The transient system is waiting for a trigger in channel 1.
The transient system is waiting for a trigger in channel 2.
The transient system is waiting for a trigger in channel 3.
0 is returned.
STATus:PRESet
command will clear all bits in the enable register.
STATus Subsystem
10
85