[Source:]Memory:current:range - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

[SOURce:]MEMory:CURRent:RANGe

Syntax
[SOURce:]MEMory:CURRent:RANGe <range>, (@<ch>)
This command stores the CURRent:RANGe <range> command into the channel
active memory list.
Parameter
Item
Type
NR1
<ch>
CPD
<range>
Remarks
This command is only applicable for the U2723A USB Modular SMU.
Example
MEM:CURR:RANG R120mA, (@1)
See Also
[SOURce:]MEMory:CURRent:LIMit
[SOURce:]MEMory:CURRent:MEASure
Keysight U2722A/U2723A Programmer's Reference
Range of values
1 through 3
– R1uA – 1 μA range
– R10uA – 10 μA range
– R100uA – 100 μA range
– R1mA – 1 mA range
– R10mA – 10 mA range
– R120mA – 120 mA range
SOURce Subsystem
Defaul t value
Required parameter
R1uA
9
61

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2723a

Table of Contents