Language; Service Menu; Date & Time; Country Setting - Sirona SiroLaser advance Operating Instructions Manual

Hide thumbs Also See for SiroLaser advance:
Table of Contents

Advertisement

 Operation
Main home screen
50
NOTICE
Please bear in mind that the system requires approx. 5 seconds to
detect the USB stick.
4. Press "OK".
 It appears the note message with the information that the
software update is in progress.
 The note message disappears, the start screen appears and the
self test is in progress.
CAUTION
Leave the USB stick and the power cable plugged in until the completion
of software update.
The Software update may take up to five minutes.
5. Please now, like when you turn on the device, press 'OK'.
 The software update was successfully performed. The USB stick
can be removed.
5.5.4.8

Service Menu

NOTICE
Only authorized persons are allowed to enter the service menu. To
avoid misuse it is necessary to enter the eight-digit pin code.

Language

5.5.4.9
With Country Settings
Language is only available if "Country Settings" are set for NON-US else
English is preset and fixed for the US (the button is grayed-out).
➢ You have the choice of twelve languages. Select one and confirm by
pressing 'OK'.
 Language will be applied after confirmation.
Date & time
5.5.4.10
Format for date: ddmmyy
Format for time (24hours notation): hhmm
➢ Enter date & time and save.

Country setting

5.5.4.11
NOTICE
The country setting for the US is pre-set.
WARNING
It's forbidden to change the country setting to Non US if you belong to
US legal regulations. The use of the country setting Non US is not
authorised by the FDA.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Advance
62 60 504 D3545
D3545.201.01.08.02 08.2012

Hide quick links: