NGL200/NGM200
Figure 6-31: File manager dialog
4. Select the file that you want to copy or delete.
5. Select the required action in the file manager dialog.
See
Table 6-6: File manager action
Action

6.14 Store and Recall

Upon power-up, the instrument loads the last stored settings from internal memory
location. Auto saved parameters are also applied during startup.
The R&S NGL/NGM output states of all channels (Ch1, Ch2) are disabled when the
recall function is activated.
Auto saved instrument settings
Auto saved of the instrument settings are applied when any of the following parameters
are changed:
Chapter 6.16, "General Instrument
USB connection mode
Ethernet settings
In addition of the auto saved instrument settings, the following instrument settings are
stored or recalled in the internal memory:
Set voltage and current level
User Manual 1178.8736.02 ─ 05
Table
6-6.
Description
Copy from internal memory to USB.
Copy from USB to internal memory.
Delete the selected file.
Settings",
on page 83
Instrument Functions
Store and Recall
74