ARBitrary:TRIGgered:MODE <mode>
ARBitrary:TRIGgered:MODE?
Sets or queries the arbitrary trigger mode of the previous selected channel.
Parameters:
<mode>
ARBitrary:TRIGgered[:STATe] <state>
ARBitrary:TRIGgered[:STATe]?
Sets or queries the arbitrary trigger mode.
Parameters:
<state>
Example:

7.7.2 Ramp

The VOTage:RAMP subsystem contains the commands for configuring the ramp func-
tion for the output channels.
[SOURce:]VOLTage:RAMP[:STATe]
[SOURce:]VOLTage:RAMP:DURation
[SOURce:]VOLTage:RAMP[:STATe] <state>
[SOURce:]VOLTage:RAMP[:STATe]?
Sets or queries the state of ramp function for the previous selected channel.
Parameters:
<state>
User Manual 1178.8736.02 ─ 05
SINGle | RUN
SINGle
A trigger event starts only with one arbitrary sequence.
RUN
A trigger event starts the whole arbitrary sequences (with all rep-
etitions).
0 | 1
0
OFF - Trigger input is deactivated.
1
ON - Trigger input is activated.
INST OUT1
ARB:TRIG ON
ARB:TRIG? -> 1
Activates the Ch1 trigger mode for arbitrary function.
................................................................................. 136
.............................................................................. 137
0 | 1
0
EasyRamp function is deactivated.
1
EasyRamp function is activated.
Remote Control Commands