NGL200/NGM200
Example:
OUTPut:IMPedance:STATe <state>
OUTPut:IMPedance:STATe?
Sets or queries the impedance target for the selected channel.
Parameters:
<state>
Example:
OUTPut:MODE <arg0>
OUTPut:MODE? <arg0>
Sets or queries the output mode.
The R&S NGL/NGM power supply series are 2 quadrant power supplies which either
source or sink current.
Parameters for setting and query:
<arg0>
User Manual 1178.8736.02 ─ 05
OUTPut:IMPedance 1
OUTPut:IMPedance? -> 1
Returns output impedance of 1 ohm.
OFF | ON | 0 | 1
OFF | 0
Deactivates output impedance for the selected channel.
ON | 1
Activates output impedance for the selected channel.
OUTPut:IMPedance 1
OUTPut:IMPedance:STAT? -> 1
Returns output impedance state as on.
AUTO | SINK | SOURce
AUTO
If operates in auto mode, the R&S NGL/NGM goes into sink or
source mode depending on the voltage across the output termi-
nal. If voltage across the output terminal exceeds the set volt-
age, current flows into the instrument, e.g. the instrument is now
operating in sink mode; vv if voltage across output terminal is
below set voltage, instrument operates as a source mode.
SINK
If operates in sink mode, current flows into the instrument. On
display, current is shown as negative current.
SOURce
If operates in source mode, current flows out from the instru-
ment.
Remote Control Commands
Configuration Commands
115