NGL200/NGM200
// selects the output channel, sets the step width
// and increases the voltage in the selected channel
// from 4 Volts
VOLT DOWN
// decreases the voltage in the selected channel
// from 4 Volts
VOLT:STEP?
// queries the voltage step size
// response: "4.000"
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Sets or queries the voltage value of the selected channel.
Parameters:
<voltage>
Example:
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement] <stepsize>
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]? [<stepsize>]
Sets or queries the incremental step size for the
Setting parameters:
<stepsize>
User Manual 1178.8736.02 ─ 05
numeric | MIN | MAX | UP | DOWN
numeric
Numeric value in V.
MIN
Minimum voltage at 0.000 V.
MAX
Maximum voltage at 20.050 V.
UP
Increases voltage by a defined step size. See
]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
on page 104.
DOWN
Decreases voltage by a defined step size. See
]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
on page 104.
Range:
0.000 to 20.05
Default unit: V
See
Example "Configuring the output voltage"
numeric | DEFault
numeric
Step value in V.
Remote Control Commands
Configuration Commands
........................................................ 104
..............................................104
[SOURce:
[SOURce:
on page 103.
VOLT UP | VOLT DOWN