Figure 6-18: Overview of trigger IO system
Table 6-3: Trigger-in signals
Trigger-in parameters
Ext trigger Ch1
Ext trigger Ch2
Digital I/O IN1
Output channel 1
Output channel 2
User Manual 1178.8736.02 ─ 05
Source
Descriptions
Digital In, pin 2 of
Digi-
If detected, corresponding trigger-out parameters
tal I/O connector
are triggered.
Digital In, pin 10 of
Dig-
The ext trigger signal is low active (inverted logic).
ital I/O connector
See
Digital In, pin 3 of
Digi-
If detected, corresponding trigger-out parameters
tal I/O connector
are triggered.
The Digital I/O IN1 signal is low active (inverted
logic).
See
Output
If respective channel output is turned on, corre-
sponding trigger-out parameters are triggered.
See
Instrument Functions
Trigger / Digital I/O
Figure
6-18.
Figure
6-18.
Figure
6-18.
59